Spectronic 20

The Spectronic 20 Colorimeter has been around for many years and we still have a few in use. It is often simply called a Spec 20. At the top of the instrument is a meter that displays both absorbance and percent transmittance of the sample being measured. On the left side is the compartment for holding the sample to be measured. On the right are the wavelength selector and indicator. On the front left is the knob that turns on the power and controls the zero position on the %T scale. On the front right is the knob that controls the 100% position on the %T scale. (These features are also diagrammed in example 8 in your workbook.) Spec 20 Colorimeter
Here are the steps to follow when using a Spec 20.

 

Turn on the instrument using the front left knob. Allow about 5 minutes for  warmup when first turned on.
Turning on the Spec 20
Select the appropriate wavelength for the sample to be measured.
Setting the wavelength on the Spec 20
With the sample holder empty and closed, adjust meter needle to 0%T  using the zero control knob.
Setting 0%T on Spec 20
Fill a tube half full with water. This is called a blank. Place it in the sample holder and close the cover.
Placing the blank sample into the Spec 20
With the blank in the sample holder and the cover closed, adjust the meter needle to 100%T using the light control knob on the right side of the instrument.
Adjusting Spec 20 to 100%T Using Blank
Remove the blank and place the sample to be measured in the sample holder and close the cover.
Placing the sample into th Spec 20
Read absorbance value (or %T) from meter. In this case the readings are 20%T and 0.70 absorbance units.
Repeat this step with additional known samples if making a calibration curve or verifying proportionality (Beer's Law).
Repeat this step with a solution of unknown concentration, so that its absorbance can be compared to the absorbance of a known solution.
Reading meter on Spec 20 with sample in place


Summary of Steps

Here is a summary of the steps of operation of a Spec 20.

General:

Specifics:

Power
Turn on power.
Warmup
Allow about 5 minutes when first turned on.
Wavelength
Select appropriate wavelength.
Zero
With sample holder empty and closed, adjust meter needle  to 0%T (or infinite A) using zero control knob.
Blank
Fill tube half full with water. Place in sample holder and close cover. Adjust meter needle to 100%T (or 0 A) using light control knob.
Standard
Measure absorbance (or %T) of known solution. Fill tube half full with sample of known concentration. Place in sample holder and close cover. Read absorbance value (or %T) from meter. Repeat this step if making a calibration curve or verifying proportionality (Beer's Law).
Sample
Measure absorbance (or %T) of solution with unknown concentration as in previous step.

 